Michael Berend contends, โFor select patients and certain indications dual mobility poly liners are a good alternative. โWeโre just not โthereโ yet with this implant, โ counters Steve MacDonald. โIn 2012 I think itโs a good option for revision total hip for instability, but not for primary total hip.โ
This weekโs Orthopaedic Crossfireยฎ debate is โThe Dual Mobility Poly Liner: The French Connection.โ For the proposition is Michael E. Berend, M.D. from the Center for Hip and Knee Surgery in Mooresville, Indiana; against the proposition is Steven J. MacDonald, M.D., F.R.C.S. (C) from the University of Western Ontario. Moderating is Thomas Thornhill, M.D. from Harvard Medical School.
Dr. Berend: โWhy select dual mobility acetabular components? You have to weigh the risks and benefits. The benefits include a proven lower dislocation rate in patients that are high risk for dislocation after total hip replacement. Those have been revision scenarios, patients who are noncompliant due to mental status changes, patients with spasticity, alcohol consumption, or patients with a history of dislocation. I think the added range of motion with dual mobility components does improve prior to impingement, both mechanical and against the pelvis. Prior concerns of polyethylene wear are largely solved with new crosslinked polys that may play a role in combination with dual mobility implants.โ
โThe risks: second decade failure mechanisms are unknown. Also, metal-metal issues on modular designs may be problematic. At least 1 in 25 patients in the Medicare database undergoing a primary total hip, and up to 14% of revision patients experience a dislocation. Hopefully this can be lowered with the use of a dual mobility component.โ
โLarge heads have helped, but dislocation continues to be a problem. I think the dual mobility implant may reduce this. It does this through increased jump distance with the larger femoral heads, and increased range of motion (ROM) prior to prosthetic or bony impingement.โ
โSo what is it? A modular acetabular component fixed to the pelvis, an uncemented or cemented stem with a smaller femoral head, and then a large polyethylene femoral head mated to the acetabular shellโฆincreasing ROM through these combined articulations.โ
โThe long term data is humbling. There are only midterm studies thus far, except for one study out to 15 years. The take home message is that dislocation is reduced with these implants compared to standard 28, 32, or 36mm implantsโฆwith equal survivorship to most clinical studies.โ
โThere are concerns about volumetric wear with increase in femoral head size. Thereโs a logarithmic increase in polyethylene wearโat least reported in the volumetric calculation. And with larger femoral heads, perhaps even to 60mm you get the increased ROM, but youโre now exposed to two interfaces, increasing the potential for volumetric wear. Hopefully the use of different articulations and polyethylenesโeither with vitamin E or crosslinking techniquesโwill reduce this in the future. With the increased ROM (at least in the lab thus far) polyethylene wear has not been increased with the combination of crosslinked poly and a dual mobility type interaction.โ
โWhat about the salvage of a well fixed metal-metal implant? Thereโs no long term data on this, but in some cases it is appealing to swap the femoral head out from a metal-metal articulation to a poly head, and avoid having to remove a well fixed acetabular component.โ
โWith new devices there are new failure mechanisms; an intraprosthetic dislocation has been reported where the small femoral head can dislocate from the polyethylene head in the cup (this is largely based on neck impingement and polyethylene wear).โ
โThere continues to be an issue of fixation with monolithic cups; some one piece monolithic cups traditionally used for metal-metal may have trouble with fixation. We have to ensure the bone stock is adequate, that thereโs not much dysplasia, and that if you donโt have the ability to use screw fixation or supplemental fixation, that fixation can be obtained. And there are concerns about metal-metal linersโฆ cobalt chrome on polyethyleneโฆinto the second decade.โ
โSo for select patients and certain indications this is a good alternative. Overall it represents about 5% of my primary practice and about 25% of my revision practice.โ
Dr. MacDonald: โThe original idea was a monoblock shell, a large polyethylene ball, and a smaller ball inside of that. In 2012 we have two options: that option with a monoblock and then a modular shell. With the latter we can get fixation with screws, metal inserts, and the same poly and metal or ceramic head.โ
โDoes this make sense? Weโre talking about a differential hardness bearing. The harder surface normally articulates against the softer surface. In most total hips done globally we do a hard (the metal/ceramic) against the soft (the poly). If you switch it around the wear rate rises exponentially.
โAnd the basic science on this construct? Zero. We have no published wear data. There is advertised data showing a 94% reduction in wearโas long as you put your shell in at 65 degrees. But if you put the shell in the normal closed position youโre not seeing that. The poly is free to go into a more closed position; that could potentially lead to edge loading and poly failure. We now have two articulating polyethylene surfaces, and we know the potential is there for increased wear and osteolysis. The claim is for improved ROM and stability, but we know from biomechanical studies that ROM of the hip peaks at about 36-38mm. And further growth in the head doesnโt improve the overall stability.โ
โWe can achieve that same ROM and stability with current implants that have registry published data. One of the constructs is the monoblock shell. The Australian registry and others show that monoblock shells have a higher failure rate than modular shells in most applications.โ
โSo who is this for? Itโs unclear. The technique manuals talk about osteoarthritic patients, rheumatoid patients, revisions, patients with a dislocation riskโฆthat could be every patient.โ
โIncreased stability is the goal, so letโs look at the dislocation rate. There is a combined paper by David Stulberg with multiple series and a dislocation rate of 0.4%. There are many studies from different centers that have dislocation rates at about 0.5 to 1% with conventional total hips. With the growth of the larger heads all the registry data show that the dislocation rate is decreasing.โ
โThere is a recent report of a new complication with this particular bearing. Itโs โintraprosthetic dislocationโโฆwith a 28mm head popping out of the polyethylene. In a large series from Remi Philippot the incidence was about 0.5%. In a recent publication from Moussa Hamadouche the incidence was 7.5% of intraprosthetic dislocations. In particular, itโs when you use a long ball that then exposes the base of the morse taper to the poly. If you look at the published clinical reports, many are from the same center, and most are short term follow up.โ
โIn a recent review article by Lachiewicz it was said, โCaution should be advised in the routine use of dual mobility components in primary and revision THA [total hip arthroplasty].โ I would echo that. They also said, โThe greatest utility may be to manage recurrent instability in revision total hip.โ There is good registry data to back this up. In a recent work by Nils Hailer there were 228 revisions for recurrent instability, and only 2% were re-revised using dual mobility.โ
โWeโre just not โthereโ yet with this implant. In 2012 I think itโs a good option for revision total hip for instability, but not for primary total hip.โ
Moderator Thornhill: โMike, you have a monoblock solid acetabulum you put inโฆthen you stick in poly and a hard surface, squeezing the poly with a smaller head. It seems Steve may be right.โ
Dr. Berend: โWe agree that for indications of trying to lessen the incidence of known problems, and perhaps accepting other problems that we donโt know. Perhaps the question is, โWill the long term complication rate be lower from the hard-hard or different mating of the materials be a lower complication rate than the known complication rate of dislocation or impingement?โโ
Moderator Thornhill: โSteve, you said you might use it in some revision situations. Which ones?โ
Dr. MacDonald: โI donโt personally use it, but perhaps when you have a recurrently unstable hipโฆlike abductor insufficiency. I use a constrained polyethylene for those scenarios, but this does make senseโฆand registry data supports its use for recurrent instability. My issue is people using it in a primary application when theyโre not paying attention to other techniques.โ
Moderator Thornhill: โMike, abductor dysfunctionโฆwill this work?โ
Dr. Berend: โIf youโre putting a fresh cup in this makes more sense for decreasing loosening of the acetabular component in the revision scenario than using constraint. I think abductor deficiency in primary total hip is probably more common than many of us think. I think the combination of that in patients with extreme high ROM in flexion, adduction, and internal rotation, with excellent acetabular bone stock. I think itโs at least an indication in my handsโespecially in smaller acetabular females with smaller acetabular bone stock.โ
Moderator Thornhill: โSo if the socket is in good position and is stable you would go to a constrained liner. If itโs loose and youโre changing to a fresh acetabular youโd be more likely to go to a new dual mobility.โ
Dr. Berend: โGreat way to think about it.โ
Moderator Thornhill: โYou okay with that, Steve?โ
Dr. MacDonald: โNo. Because constrained liners are not all created equal. Every company has a different one; some have good track records, some donโt. Some allow more ROM, others donโt. The constrained liner I use has a very good ten year clinical track record with cup loosening rates that are much lower than that intraprosthetic dislocation rate.โ
Moderator Thornhill: โSteve, what is a โlong ball?โโ
Dr. MacDonald: โAn 8 ball, 9 lengthโฆnot a 0. As you grow that longer neck then youโre exposing the base of the Morse taper to the poly.โ
Moderator Thornhill: โMike, you use this in 5% of your primariesโฆwho are they?โ
Dr. Berend: โTwo patient populations. Those with extremely high ROM (we and others have shown that preoperative ROM is a predictor of postoperative dislocation). Then itโs with the ultra-obese patient where itโs a goat rope trying to get the acetabulum in and put the screws in, get the liner in. So if I can put in a monolithic cup with good bone stock thatโs good.โ
Moderator Thornhill: โWhat about people with osteonecrosis of the femoral head?โ
Dr. Berend: โNo. I would use an anterior or anterolateral approach to preserve the posterior capsule, and try to get to at least a 36mm head.โ
Moderator Thornhill: โSteve, are our materials going to improve such that this concept will survive over the next 30 years?โ
Dr. MacDonald: โPerhaps. The weak link now is the smaller diameter ball with the poly, so if we can avoid impingement there, perhaps with second and third generation crosslinked polys we wonโt see that failure mechanism. The monoblock shell issue is dissociated from the poly issueโฆand monoblock shells in the majority of hands are not doing as well as modular shells.โ
Moderator Thornhill: โThank you to both speakers.โ